eraflash combines
maximum safety with
highest performance.
No Open Flame,
No Glowing Wire,
No Fire Hazard
The innovative Continuously Closed Cup Flash Point (CCCFP)
methods ASTM D6450 and D7094 use only 1–2 mL of sample
that is sealed in a closed chamber during the whole measurement
process. An electrical arc inside the chamber ignites the sample
and the lack of oxygen extinguishes the flame automatically. The
device measures the corresponding pressure increase inside the
chamber. An increase of ≥ 20 kPa is defined as the flash point.
High Speed
eralytics’ patented PBT – Peltier Boost TechnologyTM allows
previously unmatched heating and cooling rates. A sophisticated
mechanism safeguards the Peltier elements from high temperatures during heat-up and allows their use during cooldown even
at temperatures that would normally harm them. This cleverly
designed temperature control system allows turnaround times
that are significantly shorter than those offered by all other
flash point instruments available on the market.
Unmatched Range
PBT also facilitates the widest measurement range on the
market within a single device. The powerful Peltiers reach
-25 °C and the safeguard mechanism efficiently shields them
from temperatures up to 420 °C. The low temperature model
ERAFLASH LT even reaches -40 °C oven temperature at
significantly faster cooling rates than ever before.
Applications
ERAFLASH is the flash point testing solution for a wide variety
of industries. It is used for specifying fuels, analyzing fresh and
used lubricating oils, and bitumen testing. The small sample
volume also renders it ideal for the more expensive samples
such as flavors & fragrances, cosmetics, paints and varnishes
or even hazardous waste.